MFA0296 - BlueBox Blower | 70 hPa | O2 compatible

The air-oxygen blending in the turbine 100% O2 safe even at very low flow rates (< 2l/min), developed for maximum efficiency even in extreme cases in intensive care with dynamic capabilities up to 150 BPM for neonate applications

Specific compressor architecture with a patented motor cooling system fully independent from the patient air flow

Document(s) to download

Specifications MFA0296 - BlueBox Blower | 70 hPa | O2 compatible


APPLICATION
Hospital care, Emergency and Transport, Neonatal, Home care
MAX STATIC PRESSURE (ΔP)
70 hPa
100% OXYGEN COMPATIBILITY
Yes
MAX FREE FLOW
320 l/mn
RISE TIME*
< 86 ms
NOISE LEVEL*
< 41 dBa
TYP. POWER CONSUMPTION*
10.7 W
MOTOR COOLING MANAGEMENT
Yes - internal (independent from patient flow)
LIFETIME (L10)*
> 40 000 hours
PRESSURE RANGE (static)
< 90 hPa
TYPE
BlueBox soundproofed casing
VOLTAGE SUPPLY
24 VDC (nom) | 28 VDC (max)

Description MFA0296 - BlueBox Blower | 70 hPa | O2 compatible


The air-oxygen blending in the turbine 100% O2 safe even at very low flow rates (< 2l/min), developed for maximum efficiency even in extreme cases in intensive care  with dynamic capabilities up to 150 BPM for neonate applications.

Specific compressor architecture with a patented motor cooling system fully independent from the patient air flow.

  • Pressure range up to 65 hPa at 80 l/mn
  • Blending air-oxygen through the turbine producing a full homogeneous gas, 100% O2 safe
  • Capability to operate safely with very low flow rate (< 2l/mn)
  • Acceleration allowing to follow the breathing cycle of the patient up to 150 bpm
  • Product fully soundproofed
  • Specific compressor architecture with a patented motor cooling system fully independent from the patient air flow - 100% O2 safe
  • Reliability of the system with patient support 24/7 and 365 days

Acceleration (rise time) : < 86 ms @ 10A max - from 5 to 60 hPa @30 l/mn


Noise level : < 41 dBa @ 50 hPa - @ 1m from air inlet


Power consumption : 10.7 W @ 30 hPa @ 30 l/mn


LifeTime : > 40 000 hours - at standard operating conditions