MFA0296 - BlueBox Blower | 70 hPa | O2 compatible

The air-oxygen blending in the turbine 100% O2 safe even at very low flow rates (< 2l/min), developed for maximum efficiency even in extreme cases in intensive care with dynamic capabilities up to 150 BPM for neonate applications

Specific compressor architecture with a patented motor cooling system fully independent from the patient air flow

Document(s) to download

Specifications


APPLICATION
Hospital care, Emergency and Transport, Neonatal, Home care
MAX STATIC PRESSURE (ΔP)
70 hPa
100% OXYGEN COMPATIBILITY
Yes
MAX FREE FLOW
320 l/mn
RISE TIME*
< 86 ms
NOISE LEVEL*
< 41 dBa
TYP. POWER CONSUMPTION*
10.7 W
MOTOR COOLING MANAGEMENT
Yes - internal (independent from patient flow)
LIFETIME (L10)*
> 40 000 hours
PRESSURE RANGE (static)
< 90 hPa
TYPE
BlueBox soundproofed casing
VOLTAGE SUPPLY
24 VDC (nom) | 28 VDC (max)

The air-oxygen blending in the turbine 100% O2 safe even at very low flow rates (< 2l/min), developed for maximum efficiency even in extreme cases in intensive care  with dynamic capabilities up to 150 BPM for neonate applications.

Specific compressor architecture with a patented motor cooling system fully independent from the patient air flow.

  • Pressure range up to 65 hPa at 80 l/mn
  • Blending air-oxygen through the turbine producing a full homogeneous gas, 100% O2 safe
  • Capability to operate safely with very low flow rate (< 2l/mn)
  • Acceleration allowing to follow the breathing cycle of the patient up to 150 bpm
  • Product fully soundproofed
  • Specific compressor architecture with a patented motor cooling system fully independent from the patient air flow - 100% O2 safe
  • Reliability of the system with patient support 24/7 and 365 days

Acceleration (rise time) : < 86 ms @ 10A max - from 5 to 60 hPa @30 l/mn


Noise level : < 41 dBa @ 50 hPa - @ 1m from air inlet


Power consumption : 10.7 W @ 30 hPa @ 30 l/mn


LifeTime : > 40 000 hours - at standard operating conditions