6516e1d766564798913637.jpg
6517010b5b005347046817.jpg
686799b6e1c6f406069107.png
68679a59b7ff4362636987.png
6516e1d766564798913637.jpg
6517010b5b005347046817.jpg
686799b6e1c6f406069107.png
68679a59b7ff4362636987.png

MFA0248 - Blower w/ Dampers | 70 hPa | O2 compatible

A balanced performance-cost blower combining our patented technology for motor cooling integrated inside a specific casing with dampers

This model is designed to benefit from standard pressure performance ratio and together with easy integration thanks to the anti vibration casing

Document(s) to download

Specifications


APPLICATION
Hospital care, Emergency and Transport, Neonatal, Home care
MAX STATIC PRESSURE (ΔP)
70 hPa
MAX FREE FLOW
320 l/mn
100% OXYGEN COMPATIBILITY
Yes
RISE TIME*
< 86 ms
NOISE LEVEL*
< 52 dBa
TYP. POWER CONSUMPTION*
11.3 W
MOTOR COOLING MANAGEMENT
Yes - internal (independent from patient flow)
LIFETIME (L10)*
> 40 000 hours
TYPE
Anti-vibration casing (blower with dampers)
VOLTAGE SUPPLY
24 VDC (nom) | 28 VDC (max)
PRESSURE RANGE (static)
< 90 hPa

A balanced performance-cost blower combining our patented technology for motor cooling integrated inside a specific casing with dampers.

This model is designed to benefit from standard pressure performance ratio and together with easy integration thanks to the anti vibration casing.

  • Pressure range up to 65 hPa at 80 l/mn
  • Blending air-oxygen through the turbine producing a full homogeneous gas
  • Integration with anti-vibration casing (blower with dampers) and reduced dimensions
  • Specific compressor architecture with a patented motor cooling system fully independent from the patient air flow - 100% O2 safe
  • Reliability of the system with patient support 24/7 and 365 days

Dynamic (rise time) : < 86 ms @ 10A max - from 5 to 60 hPa @30 l/mn


Noise level : < 52 dBa @ 30 hPa -  @1m from air inlet


Power consumption : 11.3 W @ 30 hPa @ 30 l/mn


LifeTime : > 40 000 hours at standard operating conditions