MFA0248 - Blower w/ Dampers | 70 hPa | O2 compatible
Cat. : Blower with dampers
A balanced performance-cost blower combining our patented technology for motor cooling integrated inside a specific casing with dampers
This model is designed to benefit from standard pressure performance ratio and together with easy integration thanks to the anti vibration casing
Document(s) to download
APPLICATION
Hospital care, Emergency and Transport, Neonatal, Home care
MAX STATIC PRESSURE (ΔP)
70 hPa
MAX FREE FLOW
320 l/mn
RISE TIME*
< 86 ms
NOISE LEVEL*
< 52 dBa
100% OXYGEN COMPATIBILITY
Yes
TYP. POWER CONSUMPTION*
11.3 W
MOTOR COOLING MANAGEMENT
Yes - internal (independent from patient flow)
LIFETIME (L10)*
> 40 000 hours
TYPE
Anti-vibration casing (blower with dampers)
VOLTAGE SUPPLY
24 VDC (nom) | 28 VDC (max)
PRESSURE RANGE (static)
< 90 hPa
A balanced performance-cost blower combining our patented technology for motor cooling integrated inside a specific casing with dampers.
This model is designed to benefit from standard pressure performance ratio and together with easy integration thanks to the anti vibration casing.
- Pressure range up to 65 hPa at 80 l/mn
- Blending air-oxygen through the turbine producing a full homogeneous gas
- Integration with anti-vibration casing (blower with dampers) and reduced dimensions
- Specific compressor architecture with a patented motor cooling system fully independent from the patient air flow - 100% O2 safe
- Reliability of the system with patient support 24/7 and 365 days
*Dynamic (rise time) : step from 5 to 60 hPa @30 l/mn
*Noise level : @30 hPa @1m from air inlet
*Power consumption : @30 hPa @30 l/mn
*LifeTime : at standard operating condition