APPLICATION
Hospital care, Emergency and Transport, Neonatal, Home care
100% OXYGEN COMPATIBILITY
No
MAX STATIC PRESSURE (ΔP)
90 hPa
MAX FREE FLOW
320 l/mn
RISE TIME*
< 68 ms
NOISE LEVEL*
< 40 dBa
TYP. POWER CONSUMPTION*
12.5 W
MOTOR COOLING MANAGEMENT
Yes - internal (independent from patient flow)
LIFETIME (L10)*
> 40 000 hours
TYPE
BlueBox soundproofed casing
VOLTAGE SUPPLY
24 VDC (nom) | 28 VDC (max)
PRESSURE RANGE (static)
> 90 hPa
A specific patented AIRFAN blower’s architecture with motor cooling, integrated within a unique soundproof casing.
High performances, soundproof, perfectly adapted to all life support applications.
- Pressure range up to 84 hPa at 80 l/mn
- Capability to operate safely with very low flow rate (< 2l/mn)
- Acceleration allowing to follow the breathing cycle of the patient up to 200 bpm
- Product fully soundproofed
- Specific compressor architecture with a patented motor cooling system allowing to extend the lifetime
- Reliability of the system with patient support 24/7 and 365 days
*Dynamic (rise time) : step from 5 to 60 hPa @30 l/mn
*Noise level : @50 hPa @1m from air inlet
*Power consumption : @30 hPa @30 l/mn
*LifeTime : at standard operating condition